Analyzer Specs:

- (Table 1) Specifications:

Parameter Low Range ** Mid Range Full Range Full Range
High Temp(1)
High Range

RANGE

0-4% & 0-10%

0-20% 0-Inversion 0-100% 0-100% 80-100%

ACCURACY*

±0.04% for 0-4%, ±0.1% for 4-10%
OIL PHASE ONLY
±0.2% OIL PHASE ONLY ±0.5% OIL PHASE ONLY OIL PHASE ±0.5%
WATER PHASE ±1.0%
OIL PHASE ±0.5%
WATER PHASE ±1.0%
WATER PHASE ±1.0%

REPEAT-ABILITY*

±0.02%

0.1%

0.1%

OIL PHASE ±0. 1 %
WATER PHASE ±0.5%
OIL PHASE ±0. 1 %
WATER PHASE ± 0.5%
WATER PHASE ± 0.5%
Resolution*

0.01%

0.1%

0.1%

0.1%

0.1%

0.1%

-Fluid Temperature
-Extended Temp.

60'-160'F

60'-220'F(2)

60'-160'F

60'-220'F(2)
60'-160'F

60'-220'F(2)
60'-160'F

60'-220'F(2)
60'-600'F

N/A
60'-160'F

60'-220'F(2)
Salinity**

NOT A FACTOR

NOT A FACTOR

NOT A FACTOR

OIL PHASE -
NOT A FACTOR
WATER PHASE -
0.5%-8%
OIL PHASE -
NOT A FACTOR
WATER PHASE -
0.5%-8%
OIL PHASE -
NOT A FACTOR
WATER PHASE -
0.5%-8%
Electronics Ambient Temperature**

32' - 130'F

32' - 130'F 32' - 130'F 32' - 130'F 32' - 130'F 32' - 130'F
Pipe Materials

STAINLESS 316L

STAINLESS 316L

STAINLESS 316L

STAINLESS 316L

STAINLESS 316L

STAINLESS 316L

Net Oil Feature

STANDARD

STANDARD

STANDARD

STANDARD

STANDARD

STANDARD

RS-422 and
4-20mA

STANDARD

STANDARD

STANDARD

STANDARD

STANDARD

STANDARD

Alarm Relay

STANDARD

STANDARD

STANDARD

STANDARD

STANDARD

STANDARD

- * All percentages are expressed as absolute water content percentages.
- ** Other ranges available, please consult Phase Dynamics with your requirements.

 

- Physical Specifications, General:

SYSTEM: ELECTRONICS:

Power Requirements:
100 - 130 VAC or 200 - 260 VAC, 50-60 Hz: 25 wafts typical, 50 wafts at turn-on
24 VDC version is available as an option.

Rack Mount Enclosure:
7.0 H x 19.0 W x 13.0 D inches; 15 lbs.

Wall Mount Explosion-Proof Enclosure:
17.4 H x 14.0 W x 9.9 D inches; 59 lbs.; NEMA 4, 7, and 9; Class 1, Div. 1, Groups C&D (See Figure 1)

Cable:
Dedicated, 19 conductor, 22 AWG, 3 twisted pairs, 1/2" diameter. (The maximum cable distance between the measurement section and the electronics is 150 feet.)

Outputs:
Analog: field selectable (4-20 or 0-20 mA) isolated, 12 bit D-to-A
Digital: RS-422/RS-485

 

Wall Mount Fiberglass Enclosure:
16.3 H x 10.5 W x 7.9 D inches; 17 lbs.; NEMA 4X (See Figure 2)

Ambient Operating Temperature:
320 to 1300F. Consult factory for other options.

Alarm Contact Set Point:
Contact closure, rated 1 Amp, 120 VAC, field definable set point and delay (0 to 300 seconds)

System Error Relay:
Contact closure, rated 1 Amp, 120 VAC, for any system error

Flowmeter Inputs:
Magnetic pulse (3 mV to 15 V max) with field selectable factor
Current (4-20 or 0-20 mA) with field entry of flow values

MEASUREMENT:

Pressure Rating:
Up to 1,500 psig depending on process connection.

Ambient Temperature:
- 10* to 1300 F Consult factory for other options.

Temperature Compensation:
Automatic

Certification:
CSA Approved,
FM Approved

Construction:
Std: 316L SS. Designed and fabricated per ASME B31

Process Connections:
1, 2, 3, & 4 inch threaded or flanged (low range)

 

- (Table 2) Water-Cut Measurement Sections:

Normal Pipe Size
Flange Size
A
B
C
D
Net Weight
Low Range Available
Full Range Available
'Z' Config Available

1 Inch

150
300
600

39.7
39.7
39.7

5.1
5.4
5.7

13.3
13.3
13.3

24
24
24

28 lbs.
31 lbs.
31 lbs.

Yes

No

Yes

2 Inch

150
300
600

41.8
42.1
42.0

6.0
6.3
6.6

14.8
14.8
14.8

24
24
24

52 lbs.
56 lbs.
60 lbs.

Yes

Yes

Yes

3 Inch

150
300
600

43.4
43.7
43.7

7.0
7.4
7.8

15.6
15.6
15.6

24
24
24

72 lbs.
85 lbs.
85 lbs.

Yes

Yes

Yes

3 Inch
High Temp

300
600

30.5
30.5

7.6
8.0

16.4
16.4

10
10

75 lbs.
75 lbs.

No

Yes

Yes

4 Inch

150
300
600

33.2
33.7
34.1

8.4
8.9
9.8

16.7
16.7
16.7

12
12
12

120 lbs.
140 lbs.
150 lbs.

Yes

Yes

Yes