|
Preparation & Validation of
Full Range Analyzers
One fundamental purpose of sampling is to verify
online electronic instruments are providing accurate indications
of water content. The method to determine water content must be
carefully considered by observing the types of crude oil present,
the installation of the equipment (both sampling and electronic
measurement installations) and the temperature, pressure, and
level of water present.
EMULSION TYPE:
An oil and water emulsion can be one or a mix of two emulsion
phases which are called oil continuous (oil surrounding water)
and water continuous (water surrounding oil). The type of emulsion
will create different challenges in the attempt to extract a representative
sample.
A. Oil continuous emulsions: These can range
from 0% water up to 90% water in the oil continuous phase. The
light oil and NGL and condensate do not have the emulsifiers in
the liquid and therefore the water is typically in a free state
and will separate out very rapidly. In low flow conditions, this
causes severe problems in sampling. The stream must be very well
mixed with a good sample probe to extract the sample. With heavy
oils the water can exist in streamers (water continuous phase)
inside of the oil continuous emulsion. This causes extreme sampling
problems due to the non- homogeneous mixture which can also flow
at a different rate than the imbedded water phase.
B. Water continuous emulsions: In very light
oils (West Texas 40-50 API oil), NGLs and condensate, the
range of this emulsion can be from 30% water to 100% water. Opening
a valve on the main line to pull a sample tends to favor the water
phase and the sample will be biased. The higher the water cut
the more difficult it is to pull a representative sample without
mixers and pitot tube sample ports. Typical water continuous emulsion
water cut range is 65-100%.
FLOW RATES:
Flow rates will impact the sample quality. Low flow allows the
more dense liquid to favor the bottom of the process pipe. Higher
flow rates begin to mix the liquids as the turbulent flow regime
is entered. If the sample port is too close to an elbow in the
pipe, the centrifugal force will tend to separate the heavier
product and send it to the outside of the elbow. Mixing due to
a static mixer or a centrifugal pump will aid in the dispersion
of one phase in the other. These must be considered carefully
since in many cases they still do not provide good distribution
and dispersion of the water.
SAMPLE PORT DESIGN:
The key points are to have a design that will remove the sample
in a representative cross section of the main line at the same
velocity as the main line. The best will sample across the area
of the main line and place the sample into a closed chamber. Most
well testing systems open directly to atmospheric pressure when
taking the sample into an open jar. Higher velocities such as
opening a port to atmospheric pressures from the main line which
is at a higher pressure can bias the sample due to the change
in velocity at the point where the sample is being pulled. If
the port is open to the atmosphere when pulling a sample the bias
may come in due to the water slipping around the exit favoring
the oil. If the sample was pulled at the same velocity as the
flowing line, the "bow wave" effect is eliminated and
water cannot slip past. The port design should take a cross section
of the main line without creating a bias for oil or water. If
the slipstream is pulled at the center of the main line, the liquids
need to be well conditioned just upstream. A large sample port
size will create several problems. A large sample port will force
the operator to pinch down on the valve so that the oil will not
spray. If the stub between the liquids in the main line is large,
the pipe between the liquids and the valve becomes a "dead
leg" where water may start to accumulate when pulling the
sample. A recommended sample port will be a 3/8" or a maximum
of 1/ 2" tubing with a through swage fitting into a weldolet
in the main pipe that will be sampled. Figure 1 gives an example
of a 1/ 2" type of system using a 45 degree cut on the tubing
in the middle of the stream. A mark is made on the tubing on the
front side of the 45 degree cut so that after it is inserted the
technician can make certain that it is facing upstream. A static
mixer is shown just upstream to mix the fluids.

1. _3000# Threadolet
2. SS-810-1-12BT Swagelok Male Connector (2 each)
3. _Inch OD X 0.035 Wall Seamless 316 Stainless Steel Tubing
4. SS-810-1-8 Swagelok Male Connector
5. SS-18VF8-A Whitey Needle Valve
6. SS-810-1-8 Swagelok Male Connector
7. _Inch OD X 0.035 Wall Seamless 316 Stainless Steel Tubing
Figure 1. Sample Port Configuration
METHODS OF HANDLING
THE SAMPLE AFTER IT IS PULLED: Emulsions will "age"
as they cool down and time has passed from its removal from the
line. This "aging" makes it difficult to process in
the lab. If the sample is pulled using a gate type of valve or
a needle valve, these operate slowly with the opening of the valve,
the velocity will change depending on the valve opening. This
will change the apparent water cut as different operators open
the valve to different port sizes. The amount of sample pulled
should be just enough to process using the analysis method of
choice. In the case of centrifuge, this should not be more than
about 100 to 200 ml. The entire sample should be centrifuged using
the ASTM API method called out in IP359/82. Karl Fischer (KF)
should only be used for 0-5% water cuts and the sample pulled
must be processed soon after obtaining the sample. Pulling a sample
for KF is difficult due to the very small size that is processed
typically only 1ml. If the sample is left to "age"
the results will not be representative of the true water cut.
Mercaptan and some sulfurs will affect KF depending upon the chemicals
used.
LABORATORY METHODS:
Most preferred is the distillation procedure although it is the
most difficult and time consuming. Centrifuge is the typical method
for well testing applications (0-100% water cuts) but, it is often
used for 0-5% water in oil. ASTM/ API Chapter 10.4 (Field Method)
and IP359/82 (Laboratory Method) are standard test methods for
centrifuge which need to be followed to obtain reasonable results.
One overlooked issue is that of the use of water saturated toluene
as a solvent. This is of particular importance when the water cut
is lower than 0.5%. Kerosene and Stoddard solvent do not suffer
from absorption of water. Several of the centrifuge vendors sell
tubes that do not have any graduations between major marks, these
are very hard to use. The procedure does not mention using the
complete sample but, many instances this is the only way to achieve
reasonable results. In this case multiple tubes are filled and
all of the numbers are added together to obtain the average results.
Notice the repeatability and reproducibility numbers at the end
of the laboratory method standard. For 0-3% water cut the repeatability
is 0.12% and the reproducibility is 0.28%. This is not between
separate samples pulled at the same time and run separately or
for higher water cuts. The ASTM standards do not address the error
above 1% water cut for centrifuge or Dean Starke. In addition,
many cases of laboratory bias have been seen. One way to quantify
the reproducibility is to obtain a 200 to 500 ml sample from a
well mixed stream where an on line analyzer is available to make
certain that the water cut is not changing during the sample pulling.
Immediately after pulling the sample shake well and separate it
into three or four samples. Mark each with a different well name
and submit to the laboratory. The results can be very interesting
but, help to shed light on the problem.
HANDLING OF THE DATA:
Data should be plotted with the lab results on the x-axis and
the water cut analyzer on the y-axis. Be certain that the data
from the meter is read at the same time when the sample is being
pulled and that it is a fairly steady reading. If it is varying
the sample becomes the average of the analyzer readings and this
is difficult to track. Differences in DCS or PLC timing of data
collection usually means that writing down the time of day and
then looking up the analyzer reading at that time does not give
useful information. A scatter plot of the data for a given pipeline
or well vs time will help in observing the probable error in sampling
and method. Fitting a best fit line through the data will show
the offset from the analyzer to the laboratory methods. After
obtaining 10-30 points the pattern should be uniformly distributed
about the 45 degree line and have correlation between the data.
During the data taking, no adjustments should be performed on
the analyzer or these should be noted on the record so that the
plots will be meaningful. The scatter of the points will give
an indication of the deviation between the two sets of data. If
a constant offset in either direction can be seen then it is time
to place this offset into the analyzer. Constantly changing the
offset in the analyzer is not productive because this is most
likely due to the reproducibility of pulling and processing the
sample. Once the graphed data shows a clear offset this should
be entered once and more data collected. If records are kept and
each change is noted the performance of the analyzer vs time will
be seen.

EXAMPLE OF DATA:
Figure 2 shows real data from a well test where the data was obtained
and compared to a centrifuge. The middle line is the best fit
line and the two outer lines are the +/- 5% lines. This data set
has a few "outliers" which are data points that have
a problem in the data collection. Even with the data limits set
at +/-5%, the samples do not always fall within these boundries.
This graph was selected because all of the data came from one
test separator with 20 wells where the sampling was typical. Sampling
was not done from the center of the 3" pipe but instead from
the side of the wall. There was no static mixer upstream and upstream
of the sample port was a 6 foot straight run. The sample port
was too large causing the operator to pinch back on the valve
to avoid splashing out of the sample jar. The water continuous
emulsion phase shows more scatter than the oil continuous which
also states that it is harder to obtain a good sample in the water
continuous phase.
SUMMARY: Sampling
is always a very difficult problem. Flow dynamics, physical properties
of the fluids, emulsion properties, temperature, pressure, physical
pipe layout, sample method and analysis procedures all make dramatic
changes in the resultant data. Data plots vs time comparing the
instrument value vs the chosen validation method usually show
trend and provide information that cannot be obtained in any other
manner. If only a random correlation exists between the two data
sets there is something very wrong with one or both systems of
measurement.

Figure 3. Correct Installation of Sampler,
Courtesy Welker Engineering Co., Sugar Land, Texas
Figure 3 shows some of the issues in the sampling
of crude oils. Notice the oil at the bottom of the pipe, after
the mixer, after the sample loop, and before the sampler. These
problems exist in many installations with the best of intentions
to eliminate them foiled by the products characteristics
as mentioned earlier. At best sampling is an art and the water cut
analyzers are making a measurement which will reduce the difficulties
in obtaining good data. The problem is in the validation of the
water cut analyzer due to the sample and laboratory issues.
Performance Acceptance Test Procedure
The purpose of this procedure is to qualify
a water cut analyzer for use in a well test application. It is
assumed that proper sampling, laboratory and handling procedures
have been implemented.
1. Obtain a Sample:
1.1 Prepare clean sample receivers not
larger than 200 ml. Assure that the selected laboratory method
is available and appropriate for the water cut to be seen.
1.2 Upon arrival at site of sampling,
verify that the water cut analyzer is seeing a relatively constant
water cut so that sampling will be over very small changes in
water. If the analyzer display is not close to the sample point,
have another technician observe the reading while the sample is
being pulled.
1.3 Using a waste bucket, purge the sample
port line of existing liquids and then grab a sample into the
container without shutting off the flow in the sample line until
an appropriate amount of liquid has been accumulated.
1.4 Write down the information in the
attached form for the conditions at sampling. This is data for
the well under test including Stream Number, water cut, temperature,
Oil Adjust, Emulsion Phase (Oil or Wat) and salinity displayed
for the current Stream number. Oil Adjust is available by pressing
MENU four times from Home at the Phase Dynamics, Inc. analyzer.
Water Adjust should always be equal to zero.
1.5 Pull two or three samples and label
them accordingly. It would be preferred that the samples have
a different water cut. Be careful that the water cut is not changing
quickly and therefore the sample will not represent the analyzer
reading.
2. Process the Samples:
2.1 Using the ASTM recommended procedures
for centrifuge or distillation process the sample noting the results
on the line with respect to the sample number. Using Karl Fischer
for water cuts higher than 5% is not recommended. Process the
samples taken within _ hour of the collection. Longer waits may
create a more stable emulsion which will bias centrifuge results.
3. Process the Data:
3.1 Calculate the Error column on the
data sheet by taking the Laboratory water cut and subtracting
the observed water cut on the analyzer.
3.2 Determine the Average Error. At the
bottom of the Error column calculate the Average Error by summing
all of the numbers in the Error column and dividing by the number
of samples. This should only be done for samples of the same crude
type and density.
3.3 Plot the data on X-Y in Excel or
by hand with the Laboratory water cut on the X-Axis and the Analyzer
water cut on the Y-Axis.
3.4 Calculate the best fit line though
the data points using Excel or another program.
3.5 Place Error Limits around the best
fit line. The ASTM standards do not cover accuracy for water cuts
greater than 1%. Typical limits seen across many field conditions
are +/-3% in the oil continuous phase and up to +/-5% in the water
continuous phase. These errors are made up of: analyzer, sample
port, sample handling, and sample processing errors. These error
bounds are to establish a reasonable definition to look at the
sampling and analyzer issues.
4. Analyzer Calibration:
4.1 Oil continuous
emulsions require an offset to be added or subtracted to correct
for the differences between the sample values and the PDI values.
This is accomplished using the Oil Adjust. Be certain to press
Enter after selecting the appropriate offset.
4.2 Water continuous
emulsions differences between samples and PDI values are compensated
for by a change in the apparent salinity entered into the analyzer.
Water Adjust should always be equal to zero. Increasing the salinity
lowers the water cut. If the analyzer is showing "Wat"
phase and the sample is consistently showing a lower water cut
than the Phase Dynamics, Inc. analyzer, change the salinity to
reflect the difference in water cut. The salinity ONLY makes a
difference in the water continuous emulsion ("Wat" )
phase. Be certain that the analyzer is in this phase before changing
the salinity to adjust for the difference between the sample value
and the PDI value when the sample was taken. Another method to
compensate salinity for water continuous emulsions would be to
obtain a good 100% water sample under process conditions and then
performing a Standard Salinity calibration. The resultant salinity
should be the most accurate water cut obtainable.
4.3 STANDARD Salinity
Calibration Routine:
4.3.1 Make certain
that the Home Menu is showing WAT for the water continuous emulsion
phase. If this changes to OIL at any time during this procedure,
your calibration will be terminated.
4.3.2 Step through
the Menu Items until Salinity Calibration Appears on the Screen,
press ENTER, the next Menu item will be Standard Salinity Calibration,
press ENTER once again. If you pressed Select and it says Detailed
Salinity Calibration do not use this method, instead press Select
once again so that Standard Salinity calibration is displayed
then press ENTER.
4.3.3 The screen
will now show the Stream XX and Water = xx.x% and ENTER Starts
Sample. Press Enter as you either start to pull a sample or know
that the flowing stream is 100% water.
4.3.4 Wait at
least 4 seconds before pressing the ENTER again to End Sample.
If you press Enter to soon it would not have enough time to sample
the frequency.
4.3.5 The next
screen is Stream xx, Enter Water Content If you obtained a 100%
water dump during this test then use Value and Select to have
99.9% on the screen and press ENTER. If instead of obtaining a
water dump of 100% water you pulled a sample and determined the
water cut, enter this value instead of the 99.9%.
4.3.6 The system
will calculate the apparent salinity and display the New salinity
and the Old salinity and request you to press ENTER if you accept
the new salt.
4.3.7 If you pressed
ENTER in the above step, the system will place the calculated
salinity into the Menu item for this selected Stream.
This procedure should
give you the best obtainable accuracy for the water phase since
the sample was taken at pressure and temperature.
Sample Data Sheet
|
Sample ID
|
Date/
Time
|
PDI
Stream#
|
PDI
Salinity
|
PDI
Temp
|
PDI Oil
Adjust
|
PDI
Phase
|
PDI
W/C%
|
Lab
W/C%
|
Error
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
Mean Error (%)_______
|