|
Preparation & Validation of
Low Range Analyzers
One fundamental purpose of sampling is to verify
online electronic instruments are providing accurate indications
of water content. The method to determine water content must be
carefully considered by observing the types of crude oil present,
the installation of the equipment (both sampling and electronic
measurement installations) and the temperature, pressure, and
level of water present.
FLOW RATES: Flow rates will impact
the sample quality. Low flow allows the more dense liquid to favor
the bottom of the process pipe. Higher flow rates begin to mix
the liquids as the turbulent flow regime is entered. If the sample
port is too close to an elbow in the pipe, the centrifugal force
will tend to separate the heavier product and send it to the outside
of the elbow. Mixing due to a static mixer or a centrifugal pump
will aid in the dispersion of one phase in the other. These must
be considered carefully since in many cases they still do not
provide good distribution and dispersion of the water.
SAMPLE PORT DESIGN:
The key points are to have a design that will remove the sample
in a representative cross section of the main line at the same
velocity as the main line. The best will sample across the area
of the main line and place the sample into a closed chamber. Most
well testing systems open directly to atmospheric pressure when
taking the sample into an open jar. Higher velocities such as
opening a port to atmospheric pressures from the main line which
is at a higher pressure can bias the sample due to the change
in velocity at the point where the sample is being pulled. If
the port is open to the atmosphere when pulling a sample the bias
may come in due to the water slipping around the exit favoring
the oil. If the sample was pulled at the same velocity as the
flowing line, the "bow wave" effect is eliminated and
water cannot slip past. The port design should take a cross section
of the main line without creating a bias for oil or water. If
the slipstream is pulled at the center of the main line, the liquids
need to be well conditioned just upstream. A large sample port
size will create several problems. A large sample port will force
the operator to pinch down on the flow so that the oil will not
spray all over. If the stub between the liquids in the main line
is large, the pipe between the liquids and the valve becomes a
"dead leg" where water may start to accumulate when
pulling the sample. A recommended sample port will be a 3/8"
or a maximum of 1/ 2" tubing with a through swage fitting
into a weldolet in the main pipe that will be sampled. Figure
1 gives an example of a 1/ 2" type of system using a 45 degree
cut on the tubing in the middle of the stream. A mark is made
on the tubing on the front side of the 45 degree cut so that after
it is inserted the technician can make certain that it is facing
upstream. A static mixer is shown just upstream to mix the fluids.
METHODS OF HANDLING
THE SAMPLE AFTER IT IS PULLED: Emulsions will "age"
as they cool down and time has passed from its removal from the
line. This "aging" makes it difficult to process in
the lab. If the sample is pulled using a gate type of valve or
a needle valve, these operate slowly with the opening of the valve,
the velocity will change depending on the valve opening. This
will change the apparent water cut as different operators open
the valve to different port sizes. The amount of sample pulled
should be just enough to process using the analysis method of
choice. In the case of centrifuge, this should not be more than
about 100 to 200 ml. The entire sample should be centrifuged using
the ASTM API method called out in IP359/82. Karl Fischer (KF)
should only be used for 0-5% water cuts and the sample pulled
must be processed soon after obtaining the sample. Pulling a sample
for KF is difficult due to the very small size that is processed
typically only 1ml. If the sample is left to "age"
the result will not be representative of the true water cut. Mercaptan
and some sulfurs will affect KF depending upon the chemicals used.
LABORATORY METHODS:
Most preferred is the distillation procedure although it is the
most difficult and time consuming. Two problems seen in distillation
methods are in not having dry, analytical grade zxylene and not
using all of the sample obtained. Centrifuge is the typical method
for well testing applications (0-100% water cuts) but, it is often
used for 0-5% water in oil. ASTM/ API Chapter 10.4 (Field Method)
and IP359/82 (Laboratory Method) are standard test methods for
centrifuge which need to be followed to obtain reasonable results.
One overlooked issue is that of the use of water saturated toluene
as a solvent. This is of particular importance when the water cut
is lower than 0.5%. Kerosene and Stoddard solvent do not suffer
from absorption of water. Several of the centrifuge vendors sell
tubes that do not have any graduations between major marks, these
are very hard to use. The procedure does not mention using the
complete sample but, many instances this is the only way to achieve
reasonable results. In this case multiple tubes are filled and
all of the numbers are added together to obtain the average results.
Notice the repeatability and reproducibility numbers at the end
of the laboratory method standard. For 0-3% water cut the repeatability
is 0.12% and the reproducibility is 0.28%. This is not between
separate samples pulled at the same time and run separately or
for higher water cuts. In addition, many cases of laboratory bias
have been seen. One way to quantify the reproducibility is to
obtain a 200 to 500 ml sample from a well mixed stream where an
on line analyzer is available to make certain that the water cut
is not changing during the sample pulling. Immediately after pulling
the sample shake well and separate it into three or four samples.
Mark each with a different well name and submit to the laboratory.
The results can be very interesting but, help to shed light on
the problem.

1. _3000# Threadolet
2. SS-810-1-12BT Swagelok Male Connector (2 each)
3. _Inch OD X 0.035 Wall Seamless 316 Stainless Steel Tubing
4. SS-810-1-8 Swagelok Male Connector
5. SS-18VF8-A Whitey Needle Valve
6. SS-810-1-8 Swagelok Male Connector
7. _Inch OD X 0.035 Wall Seamless 316 Stainless Steel Tubing
Figure 1. Sample Port Configuration
HANDLING OF THE DATA:
Data should be plotted with the lab results on the x-axis and
the water cut analyzer on the y-axis. Be certain that the data
from the analyzer is read at the same time when the sample is
being pulled and that it is a fairly steady reading. If it is
varying the sample becomes the average of the analyzer readings
and this is difficult to track. Differences in DCS or PLC timing
of data collection usually means that writing down the time of
day and then looking up the analyzer reading at that time does
not give useful information. A scatter plot of the data for a
given pipeline or well vs time will help in observing the probable
error in sampling and method. Fitting a best fit line through
the data will show the offset from the analyzer to the laboratory
methods. After obtaining 10-30 points the pattern should be uniformly
distributed about the 45 degree line and have correlation between
the data. During the data taking, no adjustments should be performed
on the analyzer or these should be noted on the record so that
the plots will be meaningful. The scatter of the points will give
an indication of the deviation between the two sets of data. If
a constant offset in either direction can be seen then it is time
to place this offset into the analyzer. Constantly changing the
offset in the analyzer is not productive because this is most
likely due to the reproducibility of pulling and processing the
sample. Once the graphed data shows a clear offset this should
be entered once and more data collected. If records are kept and
each change is noted the performance of the analyzer vs time will
be seen.

Figure 2. Low Water Cut Data
EXAMPLE
OF DATA: Figure 2 shows real data from a pipeline application
where the data was obtained and compared to a centrifuge. The
middle dashed line is the best fit line and the two outer lines
are the +/- 0.28% lines which is the API standard for reproducibility.
Reproducibility is defined as the difference between different
operators using the same method with different equipment on the
same material. Each data point above may have been a different
operator but, with the same equipment, and different samples.
This data set has a few "outliers" which are data points
that have a problem in the data collection. Other than these few
points the reproducibility of the information appears to be close
to the +/-0.28%. In this case the mean or best fit line goes through
zero intercept and therefore the Cal Factor in the analyzer is
correct. It is easily seen from this plot that at each point if
the operator changed the Cal Factor to make the analyzer match
the lab, they would be changing the baseline without reason. If
this was done the data could not be used to determine process
variance unless each change of Cal Factor was known.
SUMMARY:
Sampling is always a very difficult problem. Flow dynamics, physical
properties of the fluids, emulsion properties, temperature, pressure,
physical pipe layout, sample method and analysis procedures all
make dramatic changes in the resultant data. Data plots vs time
comparing the instrument value vs the chosen validation method
usually show trend and provide information that cannot be obtained
in any other manner. If only a random correlation exists between
the two data sets there is something very wrong with one or both
systems of measurement.

Figure 3. Correct Installation of Sampler,
Courtesy Welker Engineering Co., Sugar Land, Texas
Figure 3 shows some of the issues in the sampling
of crude oils. Notice the oil at the bottom of the pipe, after
the mixer, after the sample loop, and before the sampler. These
problems exist in many installations with the best of intentions
to eliminate them foiled by the products characteristics
as mentioned earlier. At best sampling is an art and the water cut
analyzers are making a measurement which will reduce the difficulties
in obtaining good data. The problem is in the validation of the
water cut analyzer due to the sample and laboratory issues.
Performance Acceptance Test Procedure
The purpose of this procedure is to qualify
a water cut analyzer for use in a pipeline crude oil application.
It is assumed that proper sampling, laboratory and handling procedures
have been implemented.
1. Obtain a Sample:
1.1 Prepare clean sample receivers not
larger than 200 ml. Assure that the selected laboratory method
is available and appropriate for the water cut to be seen.
1.2 Upon arrival at site of sampling,
verify that the water cut analyzer is seeing a relatively constant
water cut so that sampling will be over very small changes in
water. If the analyzer display is not close to the sample point,
have another technician observe the reading while the sample is
being pulled.
1.3 Using a waste bucket, purge the sample
port line of existing liquids and then grab a sample into the
container without shutting off the flow in the sample line until
an appropriate amount of liquid has been accumulated.
1.4 Write down the information in the
attached form for the conditions at sampling. This is data for
water cut, temperature, Cal Factor and known density of the crude.
Cal Factor is available at the first or second menu of the Phase
Dynamics, Inc. analyzer. This is to assure that the Cal Factor
has not been changed between samples. If the analyzer reads 0.00%
water cut at any point in the data collections, an adjustment
to the Cal Factor should be made to always see a positive number
on the Display. If this is not done the calculated number could
be negative and this is not displayed on the Analyzers screen.
If a change in Cal Factor is made the data should be restarted.
1.5 Pull two or three samples and label
them accordingly. It would be preferred that the samples have
a different water cut. Be careful that the water cut is not changing
quickly and therefore the sample will not represent the analyzer
reading.
2. Process the Samples:
2.1 Using the ASTM recommended procedures
for centrifuge, KF or distillation process the sample noting the
results on the line with respect to the sample number. Process
the samples taken within _ hour of the collection.
2.2 If an on line densitometer is not
available, laboratory density can be obtained by having another
sample pulled at the same time which will be large enough to use
a hydrometer. Keep that sample separate from the water cut sample.
Recognize that this density will be different from the flowing
density obtained by an on line densitometer unless it has been
corrected to temperature. The Phase Dynamics, Inc. analyzer water cut
can be compensated in the customers PLC or DCS by using
FLOWING DENSITY for density correction. This is not density which
has been temperature corrected to 60º F.
3. Process the Data:
3.1 Calculate the Error column on the
data sheet by taking the Laboratory water cut and subtracting
the observed water cut on the analyzer.
3.2 Determine the Average Error. At the
bottom of the Error column calculate the Average Error by summing
all of the numbers in the Error column and dividing by the number
of samples. This should only be done for samples of the same crude
type and density.
3.3 Plot the data on X-Y in Excel or
by hand with the Laboratory water cut on the X-Axis and the Analyzer
water cut on the Y-Axis. USE ONLY the data from the same source
with the SAME DENSITIES for this plot. If the data encompasses
more than one crude density then the density correction routine
must be applied to the analyzer water cut (see density correction
documents).
3.4 Calculate the best fit line through
the data points using Excel or another program.
3.5 Place Error Limits around the best
fit line depending upon the laboratory method used, see Figure
2 above for an example. For Centrifuge these error limits would
be +/- 0.28%, Karl Fischer would be +/- 0.1% and Dean-Starke distillation
would be +/-0.11%. These limits are ONLY tied to the reproducibility
of a single "master" sample not between samples taken
at different times. Sample to sample variations would be much
larger than these numbers. The drawing of these error bounds is
to establish a reasonable definition to view the sampling and
laboratory issues statistically.
3.6 Observe data points that do not fall
inside of the Error limits and mark these samples as Outliers
which should NOT be included in the calculation of the Mean Error
in Section 3.2. Recalculate this Mean Error.
3.7 If the Mean Error is added to the
Phase Dynamics, Inc. water cut numbers obtained during sampling
a new graph can be generated which will be Laboratory Water Cut
on the X-Axis and Analyzer Water Cut + Mean Error on the Y-Axis.
This will show the performance of the analyzer after the Cal Factor
has been changed in the next section.
4. Analyzer Calibration:
4.1 The Mean Error can be a positive
or negative number which becomes the Calibration Factor to be
entered into the Phase Dynamics, Inc. Menu for this offset. Now
the analyzer will be reading the mean value of the results obtained
in Section 3.
4.2 If density changes the water cut
will be affected The calibration performed above should be correct
for all densities except for an offset calculated through the
equation given in the Density Correction write up from Phase Dynamics,
Inc.
Sample Data Sheet
|
Sample ID
|
Date/
Time
|
Operator
|
Flowing
Density
|
Lab
Density
|
PDI
Temp
|
PDI Cal Factor
|
PDI
W/C%
|
Lab
W/C%
|
Error
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
Mean Error (%)_______
|