|
Analyzer
Technical Information
Frequently
Asked Questions
1. How does it all work?
The electrical characteristics of the
water (68-80) and the oil (2.5) are very different and this
provides the means to clearly determine the water content. An
electrical signal is sent from the electronics on the end of
the measurement section down through the fluids. This generates
a standing wave similar to the vibrations of a rubber band held
at both ends and plucked. This standing wave changes position
within the section as the water content changes. The change
in position is automatically detected by the microwave oscillator
that originally sent out the signal and it changes its basic
frequency depending upon how much water is in the section. In
summary, the sending and receiving portion of the electronics
are the same and they change the frequency with respect to water
content.
back to top
2. Microwave measurement - the old way
Microwave energy provides for a more
complex measurement and provides enough information to improve
the measurement to include absorptive crudes. Once the absorption
becomes too great (such as in the water continuous phase), the
measurement once again degrades to a point where the resolution
of the parameters are too small to be effective.
An additional problem with standard
microwave equipment is the complexity of the equipment required
to make the measurement. A stable oscillator with amplifiers
is used to obtain the frequency and microwave energy for the
measurement. A sample must be obtained of the energy leaving
the system to compare to the energy transmitted or reflected
in the measurement section. The energy must then be transmitted
across a barrier (a seal plug or the like) into the fluids under
measurement. The introduction of this energy must be done in
accordance with electrical design constraints in order to actually
get it into the fluids instead of reflected back into the source.
This is comparable to the lens of a camera. The more expensive
camera lens has special coatings which prevent reflections and
therefore get more light through the lens and onto the film.
The energy then enters the fluids and
propagates along the structure of the measurement section which
is again designed to allow proper electrical properties to contain
and control the energy. Typically, an additional barrier (microwave
window and seal plug) and the associated structure to allow
the energy to be received must be introduced into the measurement
section. This allows the energy which has now been affected
by the fluids under measurement to be sent back to the electronics
and measured and compared to the energy originally sent out.
The measurement at this point takes
the energy incident upon (introduced into) the measurement section
from the oscillator and amplifiers and compares it to the received
energy. This is done with additional amplifiers, amplitude detectors,
phase detectors, mixers and local oscillators for the down conversion.
This is technology which is standard in many military and commercial
communication systems. The problem here is twofold; one is the
complexity of the system, and the second is that all of this
still doesn't give the required sensitivity for obtaining a
good measurement in the water phase and in other situations.
The complexity of standard microwave systems
lead to:
A) short
time between failures
B) difficulty
of maintenance (most electronic technicians cannot troubleshoot
such a system)
C) expense
of the parts
D) the
requirements for two microwave "windows" into the fluid
E) sensitivity
is still reduced from what is required.
Typically,
these systems obtain increased sensitivities by using a resonant
structure (similar to an organ pipe, the length and diameters
determine the note or resonance). The microwave energy is introduced
into the fluids and the measurement section is designed as the
resonant structure. The problem with a resonant structure is that
it goes away when you introduce absorptive materials (such as
salt water or some crudes). This is like placing a rubber plug
into the organ pipe, the strength of the vibrations is vastly
reduced and the clarity of the tone is gone.
back
to top
3. What
benefits does Oscillator Load Pull have?
What is
required to make a good measurement comes down to reduction of
the complexity and improvement of the sensitivity. Oscillator
load pull is a parameter of an oscillator which is a measure of
how much the oscillator changes its frequency when the load that
is connected to it changes. This is a parameter which is usually
designed out by preventing the oscillator from ever seeing the
changing load through the use of isolation means. Typically amplifiers
will be used after the oscillator to isolate the oscillator from
the load. This is like a guard rail around the pipe organ's pipes
to prevent people from touching the pipes and therefore changing
their tone. Interestingly, no one ever used load pull for a measurement
before, they just always knew that they needed to design it out.
Therefore, a patent was issued on this technique.
The advantages
of load pull are simplicity; the source of the microwave frequency
and energy is both the transmitter and the receiver. In addition,
the use of oscillator load pull increases the sensitivity to the
measurement parameters by 100 to 1000 times over conventional
microwave measurement techniques. This increase in sensitivity
also brings the capability to measure highly absorptive fluids
such as 28% salt in water. Because of simplicity the entire system
is very easy to troubleshoot and is very stable and reliable.
back
to top
4. What
makes Phase Dynamics different from other technologies?
The patented
measurement method which allows the oscillator to change frequency
with the changing water cut provides for up to 1000 times the
sensitivity than other technologies. This is the reason that water
continuous measurements in the high water cut region can be done.
Prior methods just did not have the sensitivity to water in the
60-90% water cut regions. This same method gives long term reproducibility
and stability of measurement because of the simplicity of the
circuitry. The engineering, materials, and methods are based on
a military style of design. Connectors, board layout, and designs
follow methods which were learned during employment with Texas
Instruments in the military electronics divisions.
back
to top
5. What
can the user check on the Measurement Section?
The only
items that can be checked on the measurement section is the tightness
of the oscillator housing, the RF front connector, and the RTD
temperature probe connections. The oscillator housing has brackets
on both sides to prevent vibration from loosening the RF Connector.
The Measurement Section has the serial number on the end metal
plug, the oscillator has a separate and different serial number
on the end of its housing. Please use the serial number on the
pipe section end when referring to the unit in correspondence,
by fax, or service call.
The microwave
electronics is a sealed box in order to prevent moisture from
the air from getting inside. The circuitry is surface mount components
because of the high frequencies involved and therefore, is not
user serviceable. The measurement section and the oscillator are
calibrated together and are considered a single unit. A unique
EPROM is created for each unit with the calibration curves generated
at the factory. If the measurement section (and electronics) has
some technical problem, the main electronics can be used with
another section if the replacement unit's EPROM is changed out
also. The measurement section has a serial number stamped on the
blank end which is also noted on the EPROM. These must match.
In addition
to the electronics, a 100 ohm RTD platinum temperature probe is
placed into the fluids just below the flange. This is connected
to the electronics through a terminal block. The Swage Lock fittings
are packed with sealant. The entire RTD and Swage system is typically
replaced as a unit. If the RTD is shorted a 100 ohm resistor can
be placed between P2 and P3 with jumpers between P+ and P2, and
another between P3 and P4 to obtain temperature while the assembly
is obtained and replaced. A suitable value may be used to obtain
a close to fluid temperature reading to maintain temperature compensation.
Installing a temporary resistor also verifies that the analog
input board is operating correctly.
back
to top
6. What
is the calibration and how is it done at the factory?
The electronics
and the measurement section are mated and placed on a fluid loop
where the calibration curves are generated. To obtain the oil
continuous phase curves (water droplets are surrounded by oil
as the continuous phase), the loop is loaded with oil and brought
to temperature. Water is injected into the recirculating oil at
a known rate with a computer system keeping track of both water
cut and the specific frequencies generated by the oscillator.
To obtain the water curves (oil droplets surrounded by water as
the continuous phase), the loop is loaded with water of a given
salinity and then oil in injected at a known rate. The loop computer
again keeps track of the water cut and the frequency of the oscillator
at each water cut. Many runs are made with various salinities
in order to fully characterize the pipe and the electronics mated
with it. This data is then curve fit into various curves of water
cut, frequency, salinity, and temperature and then placed into
EPROM as polynomial equations. This is the data set that is selected
when a salinity calibration is performed in the field. In the
field, inversion from oil continuous phase to the water continuous
phase occurs between 60% and 90% water content and depends on
the salinity, temperature, and makeup of the crude oil.
back
to top
7. What
are the considerations in mounting and use?
1) Flow
Rates and Mounting of the Measurement Section
The orientation
does not matter if the flow rate is above approximately 2-3 feet/second.
This assures a homogeneous mixture in the measurement section.
If there is gas present, the preferred orientation is with the
electronics end down and the flow from the bottom to the top.
This assures that the free gas is removed from the section instead
of accumulating it at one end. If the flow rate is very low (below
1 ft/sec), then the preferred orientation is the electronics upward
and the flow from the top to the bottom. This is so that the water
does not accumulate in the section. If the section is mounted
horizontally and the flow rate is low sand may also accumulate
in the section. Flow in the main line must be high enough to prevent
separation for a low water cut unit where a hot tap on the side
of a pipeline with a pitot tube in the upstream section is used.
The measurement section should not be at the lowest or the highest
place in a piping line unless due consideration is given to the
hold up of water, oil and gas (by maintaining high flow rates).
2) High Temperature Units
For high
temperature (>100º C fluid temperature) operation, the
electronics should be mounted football down due to the necessity
of keeping the electronics portion below 120º F ambient.
Otherwise, the heat rises and prevents the aluminum explosion
proof box from radiating the heat to the ambient air. The electronics
includes a very small built in heater to keep the electronics,
which are temperature sensitive, at a temperature of 160º
F. As the enclosure temperature (around the electronics) gets
above approximately 120º F, the heater starts to regulate
less. Above 140º F, the heater must almost shut down causing
greater temperature changes at the circuit and therefore, there
is a greater error in the measurement. There are several versions
of the Phase Dynamics measurement sections for various temperature
ranges. When the fully brazed system is selected for very high
temperatures, the electronics on the end of the measurement section
should be air purged with dry instrument air in order to maintain
the oscillator at a reasonable temperature. Contact the factory
about the requirements for air purging.
3) Electrical
Considerations
Cables and
Grounding: Included with each system are a system cable and a
green grounding wire. The system cable must be pulled from the
measurement section end back to the main electronics due to the
military type connector on one end. The green ground wire is to
meet CSA requirements and also to assure a good earth ground is
obtained between the measurement section and the electronics.
This prevents ground loops from being carried in the system cable's
shields. In all cases, the AC Input board should have an earth
ground wire connected. In addition to the safety aspects, this
assures a solid instrument ground.
Flow Meter
Input Connections: Pulsed flow inputs need to be at least 0.030
Volts and a maximum of 15 Volts. A magnetic pickup can be directly
tied in without a preamp if the cables are run in conduit and
are not excessive in length. A solid ground connection must be
made at the flow meter and at the Phase Dynamics' electronics
in order to assure extra pulses are not obtained. Current flow
inputs can be used in conjunction with other devices in the current
loop only if the Phase Dynamics is the last instrument on the
loop's path nearest to the ground leg.
back
to top
8. What
input and output options are available?
1) Digital
Communications
The RS-422
option can include a multidrop software as a purchased option
so that one RS-422 line can talk to multiple units. This must
be specified at the time of purchase since the standard unit does
not include multidrop capabilities. When the multidrop option
is implemented it becomes an RS-485. The RS-422 is implemented
with ASCII protocol which is included at the back of the user's
manual.
2) 0-20
mA and 4-20 mA Outputs
Standard
as an output is the 0-20 mA or the 4-20 mA output of water cut.
Totalization of the net oil and water can only be obtained from
the digital communications buss.
3) Well
Select OPTIONAL Input using 0-20 mA or 4-20 mA
An option
for use of the flow 4-20 mA input as a Well Select is available.
This option gives well selection up to 50 wells. This does preclude
use of a current flow input for net oil.
back
to top
9. Water
Cut measurement and salinity of the water
In order
to make any electrical measurement of water containing salts,
the salinity MUST be considered if any measurement of consequence
is done. Simply speaking, the reproducibility and accuracy of
the measurement is completely compromised without salinity compensation.
If the fluids are in the oil continuous phase - oil surrounds
the water droplets at water cuts typically less than 60% - salinity
does not affect the measurement. Salinity becomes an issue
in the water continuous phase (above about 60% water).
From the
early 1950's, it has been shown and known that measurements in
the oil-in-water emulsion phase are made difficult by the high
conductivity of the salt water. In this phase, the oil droplets
appear as voids in the conductive water and therefore, the nature
of the oil does not affect the measurement. Any change in the
salinity makes a large difference in the conductivity. This is
in parallel with the measuring capacitance of standard capacitance
and absorption probes and prevents an effective measurement from
being performed. Although some probes claim to work on attenuation
of the signal, the water is so conductive that the measurement
is very minimal at best in the water continuous phase. A good
measurement must include the effects of the salt.
The 1958
Oil and Gas Journal paper by Mr. Robert S. Wood, Capacitance-type
B.S. and W. Recorder Features Automatic, Continuous Operation,
says that when the water continuous phase is entered, the "resistance
of the emulsion has dropped so low that the probe is practically
short-circuited." He gives the resistance (one divided by the
conductivity) for various salinity waters versus water cut. It
is obvious where the water continuous phase begins. Other literature
gives the equations for the conductivity vs salt and percent water.
Another
paper from the IEEE Trans. on Microwave Theory and Techniques,
August, 1971, which gives the response for salt in water is by
A. Stogryn and is titled Equations for Calculating the Dielectric
Constant of Saline Water. Again it is obvious that the parameters
of measurement are all dependent upon both temperature and salinity
to a high degree. Salinity and temperature MUST be considered
and corrected in the measurement for accuracy and reproducibility.
back
to top
10.
Phase Dynamics salinity calibration methods
The Phase
Dynamics Water Cut Analyzer provides two means by which to enter
and determine the salinity of the produced water.
1) Use
of the Refractometer
The first
is through the use of a salinity refractometer. This is a hand
held optical instrument which uses the principle of the greater
the salinity, the greater the refraction of the room light entering
the prism area of the instrument. The user obtains a clean sample
of the well's produced water (oil in the water will cause the
line to be fuzzy), zero the refractometer, and then determine
the salinity of the produced water.
The refractometer
is zeroed by placing distilled water on the prism area and using
the adjustment screw on the top to set the blue line on zero.
The produced water should be allowed to cool to the same temperature
as the distilled water used for the calibration. This corrects
for temperature effects.
The scale
on the left gives a direct readout of percent salinity (0-10%
or 0-5%) while the right hand scale is the density in grams/cc.
This salinity value in percent can be directly entered into the
Phase Dynamics menu for salinity. This is done when a water continuous
emulsion does not exist in the well stream and a nearly correct
value is required for the future time when the well will go water
continuous.
2) Salinity
Calibration Mode of the Phase Dynamics
The preferred
method to determine and enter salinity is through the "Salinity
Calibration" which is in the Phase Dynamics' Menu. A "Standard
Salinity Calibration" allows the calculation of the salinity by
knowledge of the water cut of the well. The refractometer measurement
of salinity is used to compare the values obtained from the "Standard"
salt calibration to check both the instrument's operation and
the validity of the sample you will pull during the calibration.
Provision for the pressing of the "Enter" push button is made
while a sample is being pulled from the stream. This push button
is again pressed when the sample is complete. Once the sample
has been analyzed for water cut through distillation or
centrifuge, the resultant water cut is entered into the Phase
Dynamics unit. This allows the Phase Dynamics system to then calculate
the salinity from the frequency and temperature obtained during
the sample pull. The water cut entered by the operator then completes
the required information to calculate the apparent salinity.
The "Detailed
Salinity Calibration" is the same except that the refractometer's
reading is input and is then used to calculate a frequency offset
which would bring the same salinity as the refractometer into
use. Please use the "Standard" calibration unless otherwise requested
by the factory.
back
to top
11.
Solutions for salinity measurement problems
1) Refractometer
Solutions
If you have
problems obtaining a clean water sample please order the following
items from the Cole Palmer Chemical Company: Watman 13mm ZC Syringe
Filter, Cat. #6858-1327 pkg. of 200 (approximate cost is US$1.00
each filter). The filter part number is not in the catalog but,
it is a standard stocked item. The filters are small round items
which connect directly to the end of the syringes. Use two filters
in series if the water is very dirty and one filter cannot eliminate
the oil film. Since the filter cost is very low, please discard
after use to prevent contamination of new samples from the possibly
higher salts from the previous sample.
The refractometer
is a validity check on the operation of the system and the sample
integrity. If the two methods vary, it is typically due to the
sample's integrity.
2) Phase
Dynamics' Salinity Calibration Solutions and Guidelines
If the calculated
salt from the "Standard" salinity calibration is greater than
+\-20% of the refractometer's reading (i.e. you read 1.5% salt
on the refractometer and the "Standard" salt cal gives you a salt
of 1.0%, this is greater than -20% of 1.5%) there was a problem
with the sample, its analysis or your salt make up. In this case
the produced water may contain very little NaCl and instead may
contain larger concentrations of less conductive salts such as
Ca2Cl3.
3) Obtaining
a Valid Sample
Typically,
the manner in which the sample was obtained is the source of the
greatest error at the higher water cuts.
a) The upstream
sample port should be ahead of a mixer or have many elbows with
high flow rates to assure mixing.
b) A pitot
or quill should be installed in the sample port to obtain a sample
out of the center of the flow.
Samples
obtained from the sides or bottom of the pipe may be biased either
way depending upon flow and oil conditions. When in doubt the
best approach is to enter a salinity of approximately 10% less
than the refractometer's reading.
If the sample
port is located some distance in front or in back of the Phase
Dynamics Measurement Section a biased sample may be obtained for
two reasons:
the operator
cannot read the Phase Dynamics Display during the sampling to
assure a constant water cut reading.the process may be dynamic
and due to flow conditions the sample time frame was too short
to encompass the changes in water cut. This would lead to another
problem where the water cut is changing rapidly and therefore
an average sample would be difficult to obtain.
4) Use
the Separator to Obtain a Water Only Dump
The Phase
Dynamics' system continually displays the instantaneous water
cut at the previous entered salinity during the sample time frame.
This should be within the operator's line of sight while the sample
is pulled to observe the change in water vs time. The best opportunity
to obtain a valid sample is to raise the two phase separator's
fluid level to get a water dump which will last for 30-60 seconds.
Then a sample can be easily obtained and validated that it was
100% water during the "Enter to Start" and "Enter to End" sample
time frame.
back
to top
12.
Capacitance measurements - What is the problem?
The key
factor missing in prior technology was enough sensitivity to effectively
make a good measurement. Capacitance measurements are good enough
for 0.5% to 15% measurements as long as the medium under measurement
does not absorb much of the low frequency energy. Some crudes
do absorb energy and therefore the measurement becomes more difficult
for standard technology since this further reduces the sensitivity.
Capacitance measurement is a single factor measurement, it just
measures the amount of energy stored between two metal plates.
This limits the capability of this type of measurement.
back
to top
13.
Phase Dynamics compared to Capacitance probes
Capacitance
probes have been used for measurements of fluids for well testing
for over 40 years. These probes are used for level detection,
water content determination, and other measurements where accuracy
and reproducibility have not been an issue. The basis of measurement
is that of the ability of a material to store electrical energy.
This is typically done at low frequencies from 100 KHz to 10 MHz.
Standard off-the-shelf electronics are available to make this
measurement inexpensively. These probes are limited to approximately
30% water cut.
Measurements
requiring resolutions below 1% water in oil have been accomplished
with capacitance probes but, the lower the water content the less
the reproducibility. Long term drift has been a problem with these
instruments from the start. Since it is fundamentally an electrical
measurement at low frequencies, the length of the probe is not
critical if the flow is homogeneous. Absorptive measurements (where
the radio frequency energy is absorbed by the fluids) are extremely
compromised using capacitance probes.
Phase Dynamics
uses high frequencies where the length of the measurement section
approaches a portion of a wavelength or longer. This provides
a full cross sectional mesurement of the process fluids. The higher
frequencies of operation give greater resolution and repeatability.
The measurement is based upon a patented technology called "load
pull" that provides sensitivity improvement of 100 to 1000
times over previous technology used in microwave receivers. The
basic benefits that a microwave measurement brings to the process
analyzer are; simplicity, molecular level measurement, smaller
size, no sampling or preconditioning, and a long time before recalibration
or maintenance.
back
to top
14.
Phase Dynamics compared to Coriolis technology
These two
technologies differ in method and parameter of measurement for
determination of water content in a crude oil stream. The Coriolis
method has an advantage in the fact that it can measure both water
cut and mass flow rate.
The Coriolis
method compares the density of the emulsion to that of oil and
water. The density of water is 0.997 grams/cc at 77 degree F while
the oils density can approximately range from 0.85 to 0.95
grams/cc depending on the crude type. Through electro-mechanical
means the tubes in the Coriolis meter are set into motion and
the resultant frequency of vibration is proportional to the density
of the fluids passing through the tubes. To calculate water cut
the 60 degree F density of both the oil and the water of the particular
well in test must be known. The water density varies due to salinity
variations, while the crude density is related to the makeup of
light and heavy ends contained within the oil. The mass flow rate
is determined by the Coriolis effect between the two tubes in
motion.
Since there
is only a small difference between the densities of the water
and oil, knowledge of the densities to the 4th decimal
place for both oil and water are required. This requires a laboratory
measurement of each wells water and oil densities for calibration.
If the oil density is low, assume 0.85 grams/cc, and the water
is assumed to be 1.0 grams/cc, then the difference of 0.15 grams/cc
represents 100% change in water cut. If the wells in test are
water flood wells, salinity may change with time which affects
the density. This technology cannot determine the phase of the
fluids (water or oil continuous) since it is only a density measurement
and therefore, cannot distinguish emulsion types.
Entrained
gas will dramatically affect the density of a petroleum stream.
Since most gases have a typical density of 0.003 grams/cc and
in the case stated above 1% water cut equals 0.0015 grams/cc change,
gas has a large effect on the results. For a 1% gas content the
density is changed at 0.0% water cut by 1%*0.85 grams/cc or 0.0085
grams/cc, which represents a 0.0085/0.0015 = 5.6% water cut error.
At 5% gas content the error is up to 28%.
Phase Dynamics
technology uses the permittivity of the oil and water at 120 MHz
for reproducible water cut determination. The permittivities for
water and oil are approximately 2.4 and 80, respectively. This
large difference provides a very accurate
determination of the water cut. Unlike capacitance probes which
cannot measure in the water continuous phase, the Phase Dynamics
patented technology provides for operation from 0% to 100% water
cut and in addition, gives the operator of the field information
as to the fluid phase of the well under test. By providing the
phase information well reservoir management can be impacted both
in the wells production lifetime and profits. When a breakthrough
from an injection well occurs, a previously oil continuous well
will become water continuous. Thus, the operator will know to
take action before irreversible damage is done. Additional issues
in reservoir management can be addressed through knowing the phase
of the fluids.
Entrained
gas affects the Phase Dynamics meter much less than in the Coriolis
technology. For a 0.0% water cut, the error for 5% gas content
is only 0.16%. This is due to the large difference in the measured
parameters (permittivity of water = 80 and oil = 2.5, gas = 1.0).
For 90 100% water cut the same 5% gas content gives an
equivalent error in water cut since it looks like oil in the measurement
method for water continuous emulsions. Usually, the very high
water cuts do not have as much gas since these wells are typically
water floods.
Coriolis ^
0.0015 gm/cc = ^ 1% w/c
Phase
Dynamics ^ 2 MHz = ^ 1% w/c
back
to top
15.
The science of Oscillator Load Pull
The key
development distinguishing Phase Dynamics' Analyzer from
prior technology is the use of "Oscillator Load Pull". This microwave
parameter had been understood by defense engineers since the 1940s,
but never used as a measurement parameter. Phase Dynamics
breakthrough was precisely to use it for measurement, on-line
and in real-time, of product compositions in solutions and emulsions,
specifically of water in oil in the petroleum industry. This "Oscillator
Load Pull" technology provides for a sensitivity increase
of up to 1000 times that of other measurement methods. Phase Dynamics
achievement was recognized in 1992 by the R&D100 Award for
Innovative Technology.
An unbuffered
oscillator changes its frequency when the complex electrical impedance
at its output changes. This is known as "Oscillator Load
Pull". In Phase Dynamics Analyzer the complex impedance
is determined by the dimension of the measurement section and
the permittivity of the material. When the permittivity changes,
the complex impedance changes and the frequency of the unbuffered
oscillator changes precisely, predictably and repeatably. Permittivity
is related to the molecular dipole moment, which represents energy
storage, and the relaxation time which represents energy loss.
These are also known as relative dielectric constant and absorption
characteristics, respectively.
Microwave
permittivity depends strongly upon the molecular structure of
a material. Polar moments are created by molecular asymmetry,
ionic bond positions, or additional intrinsic details of the molecule
and its environment. This gives rise to the microwave permittivity.
Permittivity of a molecular structure will be changed by slight
position or bonding changes.
Microwave
energy will rotate polar molecules about their electrical axes
within a small microwave field. Molecules are rocked individually
or as groups which have orientational or interfacial polarizations.
It is this effect that gives rise to a change in the oscillator
frequency and gives Phase Dynamics Analyzers their excellent
sensitivity. As molecular structures are changed or as free ions
exist during the course of reaction, the microwave oscillator
exhibits a unique response which encompasses the above phenomena.
Water is
an example of a compound with a large permittivity because of
its natural polar moment. Since the two hydrogen atoms occupy
a non-symmetrical position about the oxygen, the moment about
the electrical center of the molecule is large.
This means
that water is a polar liquid. It has a relative permittivity of
approximately 80.
Benzene
is an example of a compound with a very low relative permittivity
because of its symmetrical structure. Benzene is a non- polar
liquid. It has a relative permittivity of approximately 2.
Measurements
based on permittivity allow a wide range of relative dielectric
constants and absorption characteristics. However measurements
based on capacitance are limited to materials with low absorption
characteristics.
**Note:
Although others fail to distinguish the difference, Phase
Dynamics is the sole inventor and only user of this unique method
of measurement.
back
to top
16.
Testing the effects of API gravity and of sulfur content
A study
on the effect of sulfur and a change in the API gravity on Phase
Dynamics water cut measurement is outlined below. Eight
samples of different crude oils with varying gravity and sulfur
were supplied by ARCO Pipeline for this evaluation. The gravity
varied from API 22 to API 40, and the sulfur from 0.343 to 2.17%.
The test
procedure included the following steps:
Measure
API gravity and correct for temperature.
Load
a flow loop including a Phase Dynamics with a sample of oil
and circulate.
Pull
a representative sample and determine the water cut by distillation.
Inject
at a constant rate from the initial water cut to 4% total
volume of water into the loop, while measuring the apparent
water cut with the Phase Dynamics.
Figure 1
shows the calibration factor required to zero out the water cut
change vs. API gravity and a best-fit straight line. From this
data a factor of 0.174% change in water cut offset per +1
degree API gravity can be determined. The sulfur data deviation
from a straight line vs. sulfur content appeared to be statistically
randomly scattered. It can be concluded that the sulfur content
is not a function of water cut error.
The calibration
factor mentioned above is used as an offset from a single point
Karl Fischer or distillation where the crude oil API gravity is
also known. With this single point calibration, when another crude
oil with a known API is to be measured, the correct calibration
offset can be determined for accurate readings.
For example:
A single point calibration was performed and gave a 0.25% water
cut and an API gravity of 30. If the water cut read on the Phase
Dynamics was 0.3% when the calibration sample was pulled, then
an offset of 0.05% is placed into the memory. If the next
crude type comes on line and its API is 22, then
(Cal API
New API) x (-0.174% factor) + (Cal offset) = New Offset
(30-22)
x (-0.174%) + (-0.05%) = -1.34% New Offset.
The Phase
Dynamics Analyzer can be corrected for varying API gravities once
a calibration for a given crude is performed. Other crude oils
of known API gravity are easily
corrected by entering a water cut offset using the above quotation.
Sulfur content has no effect on the measurement.
back
to top
17.
Effects of gas on Phase Dynamics
Gas carryover
in two-phase or three-phase separators can occur if the flow rate
is too high to give enough residence time, or if the separator
is partially filled with sand or silt. Another case which will
release gas from solution is where there is a pressure drop large
enough to cause flashing of the light ends. In these instances,
the entrained gas can be carried over into the measurement instrumentation,
and the effect of this free gas needs to be evaluated.
In the Phase
Dynamics Analyzer, the measurement is made using the microwave
permittivities of the gas, the oil, and the water, which are 1.0,
2.5, and 80 respectively. Since the gas and oil have low permittivity,
the gas looks like extra oil and therefore, the effect at lower
water cuts is to reduce the measured water content. As the water
percentage goes above 50% the gas effect is increased, since the
gas is replacing a volume in the fluid of which water is a great
part. As the fluid enters the water-continuous phase (typically
between 60 and 90% water content), the error in water cut due
to the gas is approximately equal to the volume content of gas.
back
to top
18.
Emulsion phase and electrical measurement
Literature
since the early 1920s describes the two physical phases of an
oil and water emulsion. One is similar to the turkey gravy
cornstarch helps to maintain a viscous water-in-oil phase. If
just water were added to the turkey oil, it would settle out into
the two discreet phases very easily. An oil-in-water emulsion
exists and is physically described as oil surrounded by water.
If we use
a coaxial line to establish the microwave energy waves with a
bare center rod and a metal outer pipe, this becomes the electromagnetic
structure to make our water cut measurement.
The oil-continuous
emulsion phase (water-in-oil) is non-conductive since the oil
is insulating. The saltwater bubbles are uniformly spaced throughout
the oil, and the salinity does not matter at the frequencies Phase
Dynamics operates (100 to 300 MHz).
When the
emulsion inverts to the water-continuous phase (oil-in-water),
then a path can be drawn from the center conductor of the coaxial
line to the outer pipe wall. Since saltwater is conductive and
the measurement is electromagnetic in nature, the
conductive liquid attenuates the electrical energy. A correction
for the conductivity must be made to know where 100% water is
on the electrical curve.
Phase Dynamics
uses the fact that when the emulsion is in the water-continuous
phase (where salinity does matter), the microwave energy is absorbed
to a much higher degree than in the oil-continuous phase. This
information is used to determine the phase of the emulsion. The
difference in attenuation of the signal between the oil-continuous
phase and the water-continuous phase is great enough that a line
of demarcation can be drawn between the two attenuation curves
for all salinities of the water.
Phase Dynamics
senses the phase of the emulsion at each measurement of 2 seconds
in time. It will respond to emulsion phase changes at any water
cut (from approximately 30% to 100% water cut). The calibration
on a flow loop is done for both emulsion phases across the entire
salinity calibration range. This is true even though the calibration
runs go only to 50% water cut for water-continuous emulsions.
The calibration runs stop at 50% water cut only due to the stability
of the water emulsion with the oil
used to make this phases calibration. Phase Dynamics has
proven that the correct phase will be selected from the attenuation
power curves generated during the water-continuous calibrations,
even for emulsions below 50%.
The Analyzer
furnished by Phase Dynamics makes a continuous emulsion phase
determination at every measurement point. This is unlike other
vendors where the emulsion phase is determined by a fixed value
of water cut to select between the curves. Only Phase Dynamicss
dynamic determination of the emulsion phase will give accurate
water cut measurements. The emulsion phase is a dynamic event,
which will change at each cycle of the well and with any change
in temperature, salinity, or oil properties. Instrumentation used
must be capable of instantaneous determination of the phase of
the emulsion for an accurate measurement.
Pulling
samples to test emulsion phase
The emulsion
phase cannot be determined easily by examination of the fluid.
Once a sample is pulled from the flowing stream, the emulsion
phase becomes unknown due to the change in temperature and flow
rate. Temperature directly determines phase because it is energy
added or substracted; flow rate is an indirect contributor.
Flow rate
contributes energy to the emulsion depending upon the Reynolds
Number of the flow region, which contributes to the shear of the
fluid. Shear adds energy as mechanical and thermal energy.
If a good
sample can be obtained, the phase is determined by a microscopic
examination of the emulsion itself, or by viscosity comparisons
to known emulsions of the same makeup. Extreme care must be taken
to assure not to disturb the emulsion type during preparation
and examination. Phase Dynamics gives an unusual insight into
the emulsion type of the fluid at a temperature and actual flowing
conditions. This information can be used to aid in the evaluation
and control of the reservoir.
back
to top
|