Analyzer Technical Information

Frequently Asked Questions

1.

How does it all work?

2.

Microwave measurement - the old way

3.

What benefits does Oscillator Load Pull have?

4.

What makes Phase Dynamics different from other technologies?

5.

What can the user check on the Measurement Section?

6.

What is the calibration and how is it done at the factory?

7.

What are the considerations in mounting and use?

8.

What input and output options are available?

9.

Water Cut measurement and salinity of the water

10.

Phase Dynamics salinity calibration methods

11.

Capacitance measurements - What is the problem?

12.

Phase Dynamics compared to Capacitance probes

13.

Phase Dynamics compared to Coriolis technology

14.

The science of Oscillator Load Pull

15.

Testing the effects of API gravity and of sulfur content

16.

Effects of gas on Phase Dynamics

17.

Emulsion phase and electrical measurement

 

 

 

1. How does it all work?

The electrical characteristics of the water (68-80) and the oil (2.5) are very different and this provides the means to clearly determine the water content. An electrical signal is sent from the electronics on the end of the measurement section down through the fluids. This generates a standing wave similar to the vibrations of a rubber band held at both ends and plucked. This standing wave changes position within the section as the water content changes. The change in position is automatically detected by the microwave oscillator that originally sent out the signal and it changes its basic frequency depending upon how much water is in the section. In summary, the sending and receiving portion of the electronics are the same and they change the frequency with respect to water content.

back to top

 

2. Microwave measurement - the old way

Microwave energy provides for a more complex measurement and provides enough information to improve the measurement to include absorptive crudes. Once the absorption becomes too great (such as in the water continuous phase), the measurement once again degrades to a point where the resolution of the parameters are too small to be effective.

An additional problem with standard microwave equipment is the complexity of the equipment required to make the measurement. A stable oscillator with amplifiers is used to obtain the frequency and microwave energy for the measurement. A sample must be obtained of the energy leaving the system to compare to the energy transmitted or reflected in the measurement section. The energy must then be transmitted across a barrier (a seal plug or the like) into the fluids under measurement. The introduction of this energy must be done in accordance with electrical design constraints in order to actually get it into the fluids instead of reflected back into the source. This is comparable to the lens of a camera. The more expensive camera lens has special coatings which prevent reflections and therefore get more light through the lens and onto the film.

The energy then enters the fluids and propagates along the structure of the measurement section which is again designed to allow proper electrical properties to contain and control the energy. Typically, an additional barrier (microwave window and seal plug) and the associated structure to allow the energy to be received must be introduced into the measurement section. This allows the energy which has now been affected by the fluids under measurement to be sent back to the electronics and measured and compared to the energy originally sent out.

The measurement at this point takes the energy incident upon (introduced into) the measurement section from the oscillator and amplifiers and compares it to the received energy. This is done with additional amplifiers, amplitude detectors, phase detectors, mixers and local oscillators for the down conversion. This is technology which is standard in many military and commercial communication systems. The problem here is twofold; one is the complexity of the system, and the second is that all of this still doesn't give the required sensitivity for obtaining a good measurement in the water phase and in other situations.

The complexity of standard microwave systems lead to:

                A) short time between failures

                B) difficulty of maintenance (most electronic technicians cannot troubleshoot such a system)

                C) expense of the parts

                D) the requirements for two microwave "windows" into the fluid

                E) sensitivity is still reduced from what is required.

Typically, these systems obtain increased sensitivities by using a resonant structure (similar to an organ pipe, the length and diameters determine the note or resonance). The microwave energy is introduced into the fluids and the measurement section is designed as the resonant structure. The problem with a resonant structure is that it goes away when you introduce absorptive materials (such as salt water or some crudes). This is like placing a rubber plug into the organ pipe, the strength of the vibrations is vastly reduced and the clarity of the tone is gone.

back to top

 

3. What benefits does Oscillator Load Pull have?

What is required to make a good measurement comes down to reduction of the complexity and improvement of the sensitivity. Oscillator load pull is a parameter of an oscillator which is a measure of how much the oscillator changes its frequency when the load that is connected to it changes. This is a parameter which is usually designed out by preventing the oscillator from ever seeing the changing load through the use of isolation means. Typically amplifiers will be used after the oscillator to isolate the oscillator from the load. This is like a guard rail around the pipe organ's pipes to prevent people from touching the pipes and therefore changing their tone. Interestingly, no one ever used load pull for a measurement before, they just always knew that they needed to design it out. Therefore, a patent was issued on this technique.

The advantages of load pull are simplicity; the source of the microwave frequency and energy is both the transmitter and the receiver. In addition, the use of oscillator load pull increases the sensitivity to the measurement parameters by 100 to 1000 times over conventional microwave measurement techniques. This increase in sensitivity also brings the capability to measure highly absorptive fluids such as 28% salt in water. Because of simplicity the entire system is very easy to troubleshoot and is very stable and reliable.

back to top

 

4. What makes Phase Dynamics different from other technologies?

The patented measurement method which allows the oscillator to change frequency with the changing water cut provides for up to 1000 times the sensitivity than other technologies. This is the reason that water continuous measurements in the high water cut region can be done. Prior methods just did not have the sensitivity to water in the 60-90% water cut regions. This same method gives long term reproducibility and stability of measurement because of the simplicity of the circuitry.

back to top

 

5. What can the user check on the Measurement Section?

The only items that can be checked on the measurement section is the tightness of the oscillator housing, the RF front connector, and the RTD temperature probe connections. The oscillator housing has brackets on both sides to prevent vibration from loosening the RF Connector. The Measurement Section has the serial number on the end metal plug, the oscillator has a separate and different serial number on the end of its housing. Please use the serial number on the pipe section end when referring to the unit in correspondence, by fax, or service call.

The microwave electronics is a sealed box in order to prevent moisture from the air from getting inside. The circuitry is surface mount components because of the high frequencies involved and therefore, is not user serviceable. The measurement section and the oscillator are calibrated together and are considered a single unit. A unique EPROM is created for each unit with the calibration curves generated at the factory. If the measurement section (and electronics) has some technical problem, the main electronics can be used with another section if the replacement unit's EPROM is changed out also. The measurement section has a serial number stamped on the blank end which is also noted on the EPROM. These must match.

In addition to the electronics, a 100 ohm RTD platinum temperature probe is placed into the fluids just below the flange. This is connected to the electronics through a terminal block. The Swage Lock fittings are packed with sealant. The entire RTD and Swage system is typically replaced as a unit. If the RTD is shorted a 100 ohm resistor can be placed between P2 and P3 with jumpers between P+ and P2, and another between P3 and P4 to obtain temperature while the assembly is obtained and replaced. A suitable value may be used to obtain a close to fluid temperature reading to maintain temperature compensation. Installing a temporary resistor also verifies that the analog input board is operating correctly.

back to top

 

6. What is the calibration and how is it done at the factory?

The electronics and the measurement section are mated and placed on a fluid loop where the calibration curves are generated. To obtain the oil continuous phase curves (water droplets are surrounded by oil as the continuous phase), the loop is loaded with oil and brought to temperature. Water is injected into the recirculating oil at a known rate with a computer system keeping track of both water cut and the specific frequencies generated by the oscillator. To obtain the water curves (oil droplets surrounded by water as the continuous phase), the loop is loaded with water of a given salinity and then oil in injected at a known rate. The loop computer again keeps track of the water cut and the frequency of the oscillator at each water cut. Many runs are made with various salinities in order to fully characterize the pipe and the electronics mated with it. This data is then curve fit into various curves of water cut, frequency, salinity, and temperature and then placed into EPROM as polynomial equations. This is the data set that is selected when a salinity calibration is performed in the field. In the field, inversion from oil continuous phase to the water continuous phase occurs between 60% and 90% water content and depends on the salinity, temperature, and makeup of the crude oil.

back to top

 

7. What are the considerations in mounting and use?

1) Flow Rates and Mounting of the Measurement Section

The orientation does not matter if the flow rate is above approximately 2-3 feet/second. This assures a homogeneous mixture in the measurement section. If there is gas present, the preferred orientation is with the electronics end down and the flow from the bottom to the top. This assures that the free gas is removed from the section instead of accumulating it at one end. If the flow rate is very low (below 1 ft/sec), then the preferred orientation is the electronics upward and the flow from the top to the bottom. This is so that the water does not accumulate in the section. If the section is mounted horizontally and the flow rate is low sand may also accumulate in the section. Flow in the main line must be high enough to prevent separation for a low water cut unit where a hot tap on the side of a pipeline with a pitot tube in the upstream section is used. The measurement section should not be at the lowest or the highest place in a piping line unless due consideration is given to the hold up of water, oil and gas (by maintaining high flow rates).

2) High Temperature Units

For high temperature (>100º C fluid temperature) operation, the electronics should be mounted football down due to the necessity of keeping the electronics portion below 120º F ambient. Otherwise, the heat rises and prevents the aluminum explosion proof box from radiating the heat to the ambient air. The electronics includes a very small built in heater to keep the electronics, which are temperature sensitive, at a temperature of 160º F. As the enclosure temperature (around the electronics) gets above approximately 120º F, the heater starts to regulate less. Above 140º F, the heater must almost shut down causing greater temperature changes at the circuit and therefore, there is a greater error in the measurement. There are several versions of the Phase Dynamics measurement sections for various temperature ranges. When the fully brazed system is selected for very high temperatures, the electronics on the end of the measurement section should be air purged with dry instrument air in order to maintain the oscillator at a reasonable temperature. Contact the factory about the requirements for air purging.

3) Electrical Considerations

Cables and Grounding: Included with each system are a system cable and a green grounding wire. The system cable must be pulled from the measurement section end back to the main electronics due to the military type connector on one end. The green ground wire is to meet CSA requirements and also to assure a good earth ground is obtained between the measurement section and the electronics. This prevents ground loops from being carried in the system cable's shields. In all cases, the AC Input board should have an earth ground wire connected. In addition to the safety aspects, this assures a solid instrument ground.

Flow Meter Input Connections: Pulsed flow inputs need to be at least 0.030 Volts and a maximum of 15 Volts. A magnetic pickup can be directly tied in without a preamp if the cables are run in conduit and are not excessive in length. A solid ground connection must be made at the flow meter and at the Phase Dynamics' electronics in order to assure extra pulses are not obtained. Current flow inputs can be used in conjunction with other devices in the current loop only if the Phase Dynamics is the last instrument on the loop's path nearest to the ground leg.

back to top

 

8. What input and output options are available?

1) Digital Communications

The RS-422 option can include a multidrop software as a purchased option so that one RS-422 line can talk to multiple units. This must be specified at the time of purchase since the standard unit does not include multidrop capabilities. When the multidrop option is implemented it becomes an RS-485. The RS-422 is implemented with ASCII protocol which is included at the back of the user's manual.

2) 0-20 mA and 4-20 mA Outputs

Standard as an output is the 0-20 mA or the 4-20 mA output of water cut. Totalization of the net oil and water can only be obtained from the digital communications buss.

3) Well Select OPTIONAL Input using 0-20 mA or 4-20 mA

An option for use of the flow 4-20 mA input as a Well Select is available. This option gives well selection up to 50 wells. This does preclude use of a current flow input for net oil.

back to top

 

9. Water Cut measurement and salinity of the water

In order to make any electrical measurement of water containing salts, the salinity MUST be considered if any measurement of consequence is done. Simply speaking, the reproducibility and accuracy of the measurement is completely compromised without salinity compensation. If the fluids are in the oil continuous phase - oil surrounds the water droplets at water cuts typically less than 60% - salinity does not affect the measurement. Salinity becomes an issue in the water continuous phase (above about 60% water).

From the early 1950's, it has been shown and known that measurements in the oil-in-water emulsion phase are made difficult by the high conductivity of the salt water. In this phase, the oil droplets appear as voids in the conductive water and therefore, the nature of the oil does not affect the measurement. Any change in the salinity makes a large difference in the conductivity. This is in parallel with the measuring capacitance of standard capacitance and absorption probes and prevents an effective measurement from being performed. Although some probes claim to work on attenuation of the signal, the water is so conductive that the measurement is very minimal at best in the water continuous phase. A good measurement must include the effects of the salt.

The 1958 Oil and Gas Journal paper by Mr. Robert S. Wood, Capacitance-type B.S. and W. Recorder Features Automatic, Continuous Operation, says that when the water continuous phase is entered, the "resistance of the emulsion has dropped so low that the probe is practically short-circuited." He gives the resistance (one divided by the conductivity) for various salinity waters versus water cut. It is obvious where the water continuous phase begins. Other literature gives the equations for the conductivity vs salt and percent water.

Another paper from the IEEE Trans. on Microwave Theory and Techniques, August, 1971, which gives the response for salt in water is by A. Stogryn and is titled Equations for Calculating the Dielectric Constant of Saline Water. Again it is obvious that the parameters of measurement are all dependent upon both temperature and salinity to a high degree. Salinity and temperature MUST be considered and corrected in the measurement for accuracy and reproducibility.

back to top

 

10. Phase Dynamics salinity calibration methods

The Phase Dynamics Water Cut Analyzer provides two manual and one automatic means by which to enter and determine the salinity of the produced water. 

 

10.1. Heuristic Salinity Routine which is a patented process by which the salinity is determined versus time.  Typically the proper salinity is selected rapidly when the fluids become water continuous.  This routine has been proven in many fields across the world and is based on statistics and the fact that a real calibration curve exists for salinity and water percentages.  This totally eliminates any operator intervention.

 

10.2. Use of the known salinity from laboratory reports.   Salinity (equivalent NaCl) given in the water reports for wells and is typically easily available.  Direct entry of the salinity will provide a reasonable beginning salt point.  The problem with this is that different salts of various species have different conductivity so it is a relative number.

 

10.3. Salinity Calibration Mode of the Phase Dynamics

The preferred method to determine and enter salinity is through the "Salinity Calibration" which is in the Phase Dynamics' Menu.  A "Standard Salinity Calibration" allows the calculation of the salinity by knowledge of the water cut of the well. The refractometer measurement of salinity is used to compare the values obtained from the "Standard" salt calibration to check both the instrument's operation and the validity of the sample you will pull during the calibration. Provision for the pressing of the "Enter" push button is made while a sample is being pulled from the stream.  This push button is again pressed when the sample is complete.  Once the sample has been analyzed for water cut through distillation or centrifuge, the resultant water cut is entered into the Phase Dynamics unit.  This allows the Phase Dynamics system to then calculate the salinity from the frequency and temperature obtained during the sample pull. The water cut entered by the operator then completes the required information to calculate the apparent salinity. 

 

The most certain method is to raise the test separator level to obtain a true 100% water dump through the analyzer.  This is easily confirmed by looking at the sample.  Only a 5-10 second dump is required to select the proper time frame to determine salinity using the Standard Salinity Calibration.

 

10.4. Obtaining a Valid Sample in the Water Continuous Phase

 

Typically, the manner in which the sample was obtained is the source of the greatest error at the higher water cuts.

 

a.) The upstream sample port should be ahead of a mixer or have many elbows with high flow rates to assure mixing. 

b.) A pitot or quill should be installed in the sample port to obtain a sample out of the center of the flow. 

 

Samples obtained from the sides or bottom of the pipe may be biased either way depending upon flow and oil conditions. 

 

If the sample port is located some distance in front or in back of the Phase Dynamics Measurement Section a biased sample may be obtained for two reasons: 1.) the operator cannot read the Phase Dynamics Display during the sampling to assure a constant water cut reading  2.) the process may be dynamic and due to flow conditions the sample time frame was too short to encompass the changes in water cut. This would lead to another problem where the water cut is changing rapidly and therefore an average sample would be difficult to obtain. 

               

The Phase Dynamics' system continually displays the instantaneous water cut at the previous entered salinity during the sample time frame.  This should be within the operator's line of sight while the sample is pulled to observe the change in water vs time. The best opportunity to obtain a valid sample is to raise the two phase separator's fluid level to get a water dump which will last for 5-20 seconds.  Then a sample can be easily obtained and validated that it was 100% water during the "Enter to Start" and "Enter to End" sample time frame.

back to top

 

11. Capacitance measurements - What is the problem?

The key factor missing in prior technology was enough sensitivity to effectively make a good measurement. Capacitance measurements are good enough for 0.5% to 15% measurements as long as the medium under measurement does not absorb much of the low frequency energy. Some crudes do absorb energy and therefore the measurement becomes more difficult for standard technology since this further reduces the sensitivity. Capacitance measurement is a single factor measurement, it just measures the amount of energy stored between two metal plates. This limits the capability of this type of measurement.

back to top

 

12. Phase Dynamics compared to Capacitance probes

Capacitance probes have been used for measurements of fluids for well testing for over 40 years. These probes are used for level detection, water content determination, and other measurements where accuracy and reproducibility have not been an issue. The basis of measurement is that of the ability of a material to store electrical energy. This is typically done at low frequencies from 100 KHz to 10 MHz. Standard off-the-shelf electronics are available to make this measurement inexpensively. These probes are limited to approximately 30% water cut.

Measurements requiring resolutions below 1% water in oil have been accomplished with capacitance probes but, the lower the water content the less the reproducibility. Long term drift has been a problem with these instruments from the start. Since it is fundamentally an electrical measurement at low frequencies, the length of the probe is not critical if the flow is homogeneous. Absorptive measurements (where the radio frequency energy is absorbed by the fluids) are extremely compromised using capacitance probes.

Phase Dynamics uses high frequencies where the length of the measurement section approaches a portion of a wavelength or longer. This provides a full cross sectional mesurement of the process fluids. The higher frequencies of operation give greater resolution and repeatability. The measurement is based upon a patented technology called "load pull" that provides sensitivity improvement of 100 to 1000 times over previous technology used in microwave receivers. The basic benefits that a microwave measurement brings to the process analyzer are; simplicity, molecular level measurement, smaller size, no sampling or preconditioning, and a long time before recalibration or maintenance.

back to top

 

13. Phase Dynamics compared to Coriolis technology

These two technologies differ in method and parameter of measurement for determination of water content in a crude oil stream. The Coriolis method has an advantage in the fact that it can measure both water cut and mass flow rate.

The Coriolis method compares the density of the emulsion to that of oil and water. The density of water is 0.997 grams/cc at 77 degree F while the oil’s density can approximately range from 0.85 to 0.95 grams/cc depending on the crude type. Through electro-mechanical means the tubes in the Coriolis meter are set into motion and the resultant frequency of vibration is proportional to the density of the fluids passing through the tubes. To calculate water cut the 60 degree F density of both the oil and the water of the particular well in test must be known. The water density varies due to salinity variations, while the crude density is related to the makeup of light and heavy ends contained within the oil. The mass flow rate is determined by the Coriolis effect between the two tubes in motion.

Since there is only a small difference between the densities of the water and oil, knowledge of the densities to the 4th decimal place for both oil and water are required. This requires a laboratory measurement of each well’s water and oil densities for calibration. If the oil density is low, assume 0.85 grams/cc, and the water is assumed to be 1.0 grams/cc, then the difference of 0.15 grams/cc represents 100% change in water cut. If the wells in test are water flood wells, salinity may change with time which affects the density. This technology cannot determine the phase of the fluids (water or oil continuous) since it is only a density measurement and therefore, cannot distinguish emulsion types.

Entrained gas will dramatically affect the density of a petroleum stream. Since most gases have a typical density of 0.003 grams/cc and in the case stated above 1% water cut equals 0.0015 grams/cc change, gas has a large effect on the results. For a 1% gas content the density is changed at 0.0% water cut by 1%*0.85 grams/cc or 0.0085 grams/cc, which represents a 0.0085/0.0015 = 5.6% water cut error. At 5% gas content the error is up to 28%.

Phase Dynamics’ technology uses the permittivity of the oil and water at 120 MHz for reproducible water cut determination. The permittivities for water and oil are approximately 2.4 and 80, respectively. This large difference provides a very accurate determination of the water cut. Unlike capacitance probes which cannot measure in the water continuous phase, the Phase Dynamics patented technology provides for operation from 0% to 100% water cut and in addition, gives the operator of the field information as to the fluid phase of the well under test. By providing the phase information well reservoir management can be impacted both in the well’s production lifetime and profits. When a breakthrough from an injection well occurs, a previously oil continuous well will become water continuous. Thus, the operator will know to take action before irreversible damage is done. Additional issues in reservoir management can be addressed through knowing the phase of the fluids.

Entrained gas affects the Phase Dynamics meter much less than in the Coriolis technology. For a 0.0% water cut, the error for 5% gas content is only 0.16%. This is due to the large difference in the measured parameters (permittivity of water = 80 and oil = 2.5, gas = 1.0). For 90 — 100% water cut the same 5% gas content gives an equivalent error in water cut since it looks like oil in the measurement method for water continuous emulsions. Usually, the very high water cuts do not have as much gas since these wells are typically water floods.

Coriolis                  ^ 0.0015 gm/cc = ^ 1% w/c

Phase Dynamics ^ 2 MHz = ^ 1% w/c

back to top

 

14. The science of Oscillator Load Pull

The key development distinguishing Phase Dynamics' Analyzer from prior technology is the use of "Oscillator Load Pull". This microwave parameter had been understood by defense engineers since the 1940s, but never used as a measurement parameter. Phase Dynamics’ breakthrough was precisely to use it for measurement, on-line and in real-time, of product compositions in solutions and emulsions, specifically of water in oil in the petroleum industry. This "Oscillator Load Pull" technology provides for a sensitivity increase of up to 1000 times that of other measurement methods. Phase Dynamics’ achievement was recognized in 1992 by the R&D100 Award for Innovative Technology.

An unbuffered oscillator changes its frequency when the complex electrical impedance at its output changes. This is known as "Oscillator Load Pull". In Phase Dynamics’ Analyzer the complex impedance is determined by the dimension of the measurement section and the permittivity of the material. When the permittivity changes, the complex impedance changes and the frequency of the unbuffered oscillator changes precisely, predictably and repeatably. Permittivity is related to the molecular dipole moment, which represents energy storage, and the relaxation time which represents energy loss. These are also known as relative dielectric constant and absorption characteristics, respectively.

Microwave permittivity depends strongly upon the molecular structure of a material. Polar moments are created by molecular asymmetry, ionic bond positions, or additional intrinsic details of the molecule and its environment. This gives rise to the microwave permittivity. Permittivity of a molecular structure will be changed by slight position or bonding changes.

Microwave energy will rotate polar molecules about their electrical axes within a small microwave field. Molecules are rocked individually or as groups which have orientational or interfacial polarizations. It is this effect that gives rise to a change in the oscillator frequency and gives Phase Dynamics’ Analyzers their excellent sensitivity. As molecular structures are changed or as free ions exist during the course of reaction, the microwave oscillator exhibits a unique response which encompasses the above phenomena.

Water is an example of a compound with a large permittivity because of its natural polar moment. Since the two hydrogen atoms occupy a non-symmetrical position about the oxygen, the moment about the electrical center of the molecule is large.

This means that water is a polar liquid. It has a relative permittivity of approximately 80.

Benzene is an example of a compound with a very low relative permittivity because of its symmetrical structure. Benzene is a non- polar liquid. It has a relative permittivity of approximately 2.

Measurements based on permittivity allow a wide range of relative dielectric constants and absorption characteristics. However measurements based on capacitance are limited to materials with low absorption characteristics.

**Note: Although others fail to distinguish the difference, Phase Dynamics is the sole inventor and only user of this unique method of measurement.

back to top

 

15. Testing the effects of API gravity and of sulfur content

A study on the effect of sulfur and a change in the API gravity on Phase Dynamic’s water cut measurement is outlined below. Eight samples of different crude oils with varying gravity and sulfur were supplied by ARCO Pipeline for this evaluation. The gravity varied from API 22 to API 40, and the sulfur from 0.343 to 2.17%.

The test procedure included the following steps:

Measure API gravity and correct for temperature.

Load a flow loop including a Phase Dynamics with a sample of oil and circulate.

Pull a representative sample and determine the water cut by distillation.

Inject at a constant rate from the initial water cut to 4% total volume of water into the loop, while measuring the apparent water cut with the Phase Dynamics.

Figure 1 shows the calibration factor required to zero out the water cut change vs. API gravity and a best-fit straight line. From this data a factor of —0.174% change in water cut offset per +1 degree API gravity can be determined. The sulfur data deviation from a straight line vs. sulfur content appeared to be statistically randomly scattered. It can be concluded that the sulfur content is not a function of water cut error.

The calibration factor mentioned above is used as an offset from a single point Karl Fischer or distillation where the crude oil API gravity is also known. With this single point calibration, when another crude oil with a known API is to be measured, the correct calibration offset can be determined for accurate readings.

For example: A single point calibration was performed and gave a 0.25% water cut and an API gravity of 30. If the water cut read on the Phase Dynamics was 0.3% when the calibration sample was pulled, then an offset of —0.05% is placed into the memory. If the next crude type comes on line and its API is 22, then

(Cal API — New API) x (-0.174% factor) + (Cal offset) = New Offset

(30-22) x (-0.174%) + (-0.05%) = -1.34% New Offset.

The Phase Dynamics Analyzer can be corrected for varying API gravities once a calibration for a given crude is performed. Other crude oils of known API gravity are easily corrected by entering a water cut offset using the above quotation. Sulfur content has no effect on the measurement.

back to top

 

16. Effects of gas on Phase Dynamics

Gas carryover in two-phase or three-phase separators can occur if the flow rate is too high to give enough residence time, or if the separator is partially filled with sand or silt. Another case which will release gas from solution is where there is a pressure drop large enough to cause flashing of the light ends. In these instances, the entrained gas can be carried over into the measurement instrumentation, and the effect of this free gas needs to be evaluated.

In the Phase Dynamics Analyzer, the measurement is made using the microwave permittivities of the gas, the oil, and the water, which are 1.0, 2.5, and 80 respectively. Since the gas and oil have low permittivity, the gas looks like extra oil and therefore, the effect at lower water cuts is to reduce the measured water content. As the water percentage goes above 50% the gas effect is increased, since the gas is replacing a volume in the fluid of which water is a great part. As the fluid enters the water-continuous phase (typically between 60 and 90% water content), the error in water cut due to the gas is approximately equal to the volume content of gas.

back to top

 

17. Emulsion phase and electrical measurement

Literature since the early 1920s describes the two physical phases of an oil and water emulsion. One is similar to the turkey gravy — cornstarch helps to maintain a viscous water-in-oil phase. If just water were added to the turkey oil, it would settle out into the two discreet phases very easily. An oil-in-water emulsion exists and is physically described as oil surrounded by water.

If we use a coaxial line to establish the microwave energy waves with a bare center rod and a metal outer pipe, this becomes the electromagnetic structure to make our water cut measurement.

The oil-continuous emulsion phase (water-in-oil) is non-conductive since the oil is insulating. The saltwater bubbles are uniformly spaced throughout the oil, and the salinity does not matter at the frequencies Phase Dynamics operates (100 to 300 MHz).

When the emulsion inverts to the water-continuous phase (oil-in-water), then a path can be drawn from the center conductor of the coaxial line to the outer pipe wall. Since saltwater is conductive and the measurement is electromagnetic in nature, the conductive liquid attenuates the electrical energy. A correction for the conductivity must be made to know where 100% water is on the electrical curve.

Phase Dynamics uses the fact that when the emulsion is in the water-continuous phase (where salinity does matter), the microwave energy is absorbed to a much higher degree than in the oil-continuous phase. This information is used to determine the phase of the emulsion. The difference in attenuation of the signal between the oil-continuous phase and the water-continuous phase is great enough that a line of demarcation can be drawn between the two attenuation curves for all salinities of the water.

Phase Dynamics senses the phase of the emulsion at each measurement of 2 seconds in time. It will respond to emulsion phase changes at any water cut (from approximately 30% to 100% water cut). The calibration on a flow loop is done for both emulsion phases across the entire salinity calibration range. This is true even though the calibration runs go only to 50% water cut for water-continuous emulsions. The calibration runs stop at 50% water cut only due to the stability of the water emulsion with the oil used to make this phase’s calibration. Phase Dynamics has proven that the correct phase will be selected from the attenuation power curves generated during the water-continuous calibrations, even for emulsions below 50%.

The Analyzer furnished by Phase Dynamics makes a continuous emulsion phase determination at every measurement point. This is unlike other vendors where the emulsion phase is determined by a fixed value of water cut to select between the curves. Only Phase Dynamics’s dynamic determination of the emulsion phase will give accurate water cut measurements. The emulsion phase is a dynamic event, which will change at each cycle of the well and with any change in temperature, salinity, or oil properties. Instrumentation used must be capable of instantaneous determination of the phase of the emulsion for an accurate measurement.

Pulling samples to test emulsion phase

The emulsion phase cannot be determined easily by examination of the fluid. Once a sample is pulled from the flowing stream, the emulsion phase becomes unknown due to the change in temperature and flow rate. Temperature directly determines phase because it is energy added or substracted; flow rate is an indirect contributor.

Flow rate contributes energy to the emulsion depending upon the Reynolds Number of the flow region, which contributes to the shear of the fluid. Shear adds energy as mechanical and thermal energy.

If a good sample can be obtained, the phase is determined by a microscopic examination of the emulsion itself, or by viscosity comparisons to known emulsions of the same makeup. Extreme care must be taken to assure not to disturb the emulsion type during preparation and examination. Phase Dynamics gives an unusual insight into the emulsion type of the fluid at a temperature and actual flowing conditions. This information can be used to aid in the evaluation and control of the reservoir.

back to top